SKU: 190002
The Central System is used during deliveries and surgeries, as well as for other applications involving gases, where multiple exhaust points are connected.
The Medicvent Central System Double is designed to handle a flow of 560 m³. For example, it can simultaneously extract 35 m³ per hour from 16 suction outlets for Medicvent Double Masks.
Note! The system does not meet the requirements of SS-ISO 16571:2024 for the evacuation of smoke from surgical instruments. According to the smoke evacuation standard, a redundancy system is required for smoke evacuation.
The system consists of two side channel fans with specific performance characteristics in terms of pressure curve and noise. It is controlled by a frequency-type motor controller programmed to minimize power losses in the fan. The system is fully automated. The motor control is connected to a Medicvent pressure regulator. The pressure regulator is attached to the fan’s suction duct and controls the motor controller to adjust the fan speed according to the required flow.
Power:
400-460VAC, 50-60Hz 3-phase +N +PE
Automatic fuse (control cabinet)
3P C10A
Power consumption:
4kW
Air flow:
0 – 560 m³/hour
Sound level fan:
73dBA
How can we help you? Contact us, and we’ll be happy to provide more information.