Medicvent Central System Single 360 is used for the evacuation of nitrous oxide and anesthetic gases. The system does not meet the requirements of SS-ISO 16571:2024 for the evacuation of smoke from surgical instruments.

The Medicvent Central System Single is designed to handle a flow of 280 m³. For example, it can simultaneously extract 35 m³ per hour from 8 suction outlets for Medicvent Double Masks.

Note! The system does not meet the requirements of SS-ISO 16571:2024 for the evacuation of smoke from surgical instruments. According to the smoke evacuation standard, a redundant system is required for smoke evacuation. Read more about our redundancy systems here.

The system consists of a side channel fan with specific performance characteristics in terms of pressure curve and noise. It is controlled by a frequency-type motor controller programmed to minimize power losses in the fan. The system is fully automated. The motor control is connected to a Medicvent pressure regulator. The pressure regulator is attached to the fan’s suction duct and controls the motor controller to adjust the fan speed according to the required flow.

Technical data

Power:

400-460VAC, 50-60Hz 3-phase +N +PE

Automatic fuse (control cabinet)

3P C6A

Power consumption:

2,2kW

Air flow:

0 – 280 m³/hour

Sound level fan:

70dBA